New auto-weaning ventilator might make pulmonologists obsolete Part 1 (RCT, AJRCCM)



Maybe it’s the surgeons who are tired of consulting us for “vent management,” and finding out we can’t get patients off the vent any faster than they can without our help. Anyway, some troublemakers in Germany (Dirk Schadler et al) found that among 300 patients in surgical intensive care units (SICUs), an automatically-weaning ventilator was as good as human beings using a ventilator weaning protocol at liberating patients from mechanical ventilation.



What They Did
Authors randomized 300 post-surgical patients at 3 SICUs at one hospital in Kiel, Germany after they had been on 9 hours of mechanical ventilation to one of two weaning strategies. Half were managed with standard ventilator-weaning protocols, managed by physicians and respiratory therapists. Half had a switch on their ventilators (Evita XL) flipped to enable the SmartCare software embedded into it. This software automatically adjusts the pressure support delivered in response to patient effort, aiming to keep a patient in a “comfort zone” of 15-30 breaths per minute, with expired paCO2 of < 55 mm Hg, and tidal volumes greater than 300 mL. PEEP had to be set and changed manually.
When the patient is staying in the respiratory comfort zone with a minimum of pressure support and PEEP, the clever ventilator automatically initiates a spontaneous breathing trial. If the patient passes the machine-initiated and monitored SBT, the ventilator goes “DING!“, then extubates the patient and transfers them out of the ICU (I’m making up this last part…I think).
The study’s primary endpoint was overall ventilation time in the ICU, including noninvasive ventilation (NIPPV/BiPAP), during the 28-day study period. Secondary end points were time in the respiratory “comfort zone”, numbers of alarms and vent changes, length of stay in the ICU and hospital, and mortality at 28- and 90-days.
They predefined three subgroups: cardiac surgery patients (n=132), patients with sepsis (n=44), and patients with COPD (n=41).
The study was funded in large part by a grant of 70,000 euros from Drager, the company that makes the SmartCare software.
What They Found
Overall, patients were weaned/liberated from mechanical ventilation equally quickly in both groups: a median of 31 hours in the automatic-weaning group, and 39 hours in the humans-using-weaning-protocols group (p=0.18).
In the predefined cardiac surgery subgroup (n=132), patients on the automatic-weaning vent were liberated a bit faster: 24 hours of ventilation, vs. 35 hours for controls.
In the sepsis and COPD subgroups, there was no statistically significant difference in ventilator hours, but numerical results were widely discordant:
  • 44 patients with sepsis “auto-weaned” in a median 101 hours (5 days) vs. 144 hours (7 days) for the human-weaning group. (p=0.296)
  • 41 patients with COPD took longer to escape the ventilator on auto-weaning than with standard human-driven protocols (55 hours vs. 28 hours, p=0.095).
There were abundant protocol violations: 25% of the time the automatic weaning software was supposed to be “on,” it was not. This would seem to make it harder to show a benefit (i.e., reduce the observed effect of auto-weaning).
Patients in the auto-wean group were placed on their first spontaneous breathing trial (SBT) by the machine within one hour in a majority of cases; those managed by humans took a median of 9 hours to start an SBT. There was a “trend” toward fewer tracheostomies in the automatic-weaning group (17 vs. 28, p=0.08).

Comments

Popular posts from this blog

Driving Pressure in ARDS: A new concept!

Driving Pressure and Survival in the Acute Respiratory Distress Syndrome Marcelo B.P. Amato, M.D., Maureen O. Meade, M.D., Arthur S. Slutsky, M.D., Laurent Brochard, M.D., Eduardo L.V. Costa, M.D., David A. Schoenfeld, Ph.D., Thomas E. Stewart, M.D., Matthias Briel, M.D., Daniel Talmor, M.D., M.P.H., Alain Mercat, M.D., Jean-Christophe M. Richard, M.D., Carlos R.R. Carvalho, M.D., and Roy G. Brower, M.D. N Engl J Med 2015; 372:747-755 February 19, 2015 DOI: 10.1056/NEJMsa1410639 BACKGROUND Mechanical-ventilation strategies that use lower end-inspiratory (plateau) airway pressures, lower tidal volumes (V T ), and higher positive end-expiratory pressures (PEEPs) can improve survival in patients with the acute respiratory distress syndrome (ARDS), but the relative importance of each of these components is uncertain. Because respiratory-system compliance (C RS ) is strongly related to the volume of aerated remaining functional lung during disease (termed functional lung size)...

Anaphylaxis updates part 2- Empty Ventricle Syndrome

Patients with anaphylaxis should not suddenly sit, stand, or be placed in the upright position. Instead, they should be placed on the back with their lower extremities elevated or, if they are experiencing respiratory distress or vomiting, they should be placed in a position of comfort with their lower extremities elevated. This accomplishes 2 therapeutic goals: 1) preservation of fluid in the circulation (the central vascular compartment), an important step in managing distributive shock; and 2) prevention of the empty vena cava/empty ventricle syndrome, which can occur within seconds when patients with anaphylaxis suddenly assume or are placed in an upright position. Patients with this syndrome are at high risk for sudden death. They are unlikely to respond to epinephrine regardless of route of administration, because it does not reach the heart and therefore cannot be circulated throughout the body

Epidural catheter tests...not only the test dose

Siphon test The catheter is held upright and a fluid level sought. If the catheter is then elevated, the fluid level should fall (see inset) as the fluid siphons in to the epidural space, which is usually under negative pressure compared with atmospheric. If the fluid column continues to rise, this may suggest subarachnoid placement. The siphon test can be reassuring, but is not mandatory. Aspiration  This should be considered mandatory. The Luer connector is attached to the catheter and a syringe is used to apply negative pressure. Free and continued aspiration of clear fluid can indicate subarachnoid placement of the catheter. However, if saline has been used for loss of resistance, it is not unusual for a small amount of this to be aspirated. If there is doubt, the aspirated fluid can be tested for glucose (cerebrospinal fluid will generally test positive) or mixed with thiopentone (cerebrospinal fluid forms a precipitate). If blood is freely and continuously aspirated, this sug...